3 inhibit sampler’s program, Inhibit sampler’s program – Campbell Scientific 10164-L Water Sampler Control Cable for use with Isco and Sigma Autosamplers User Manual
Page 15

10164-L Sampler Control Cable for use with Isco and Sigma Autosamplers
3: Do (P86)
1: 10
Set high Flag 0 (output flag)
4: Set Active Storage Area (P80)
1: 1
Final Storage Area 1
2: 400
Array ID or location
5: Real Time (P77)
1: 1110
Year,Day,Hour-Minute
6: Sample (P70)
1: 1
Reps
2: 10
Loc [ Sample_No ]
7: End (P95)
;of Subroutine Number 98
End Table 3
4.2.3 Inhibit Sampler’s Program
Wire Color
CR10(X)
Purple
G
(Isco) Green
C1
(Sigma) Brown C1
Clear
G
To inhibit an Isco sampler from running its own program, set control port 1
low using a program control instruction such as Instruction 86. To allow the
sampler to run its program, set it high.
The logic for Sigma samplers is just the opposite. A high signal
inhibits the sampler.
Example for inhibiting an Isco sampler's program:
Do (P86)
1: 51
Set Port 1 low
; (41 would set it high)
NOTE