CS420-L and CS425-L Pressure Transducers
7.2.2 Instruction 9 (6-Wire Full Bridge w/Excitation)
7.2.2.1 Wiring
Datalogger
CS420-L/CS425-L
Excitation
E1
Red (Supply +)
Input
H1
Orange (Monitor +)
Input
L1
Black (Monitor -)
Input
H2
Yellow (Output +)
Input
L2
Blue (Output -)
G
White (Supply -)
G
Clear
(Shield)
7.2.2.2 Edlog Program
; Measure one 6-wire DRUCK Sensor.
* Table 1 Program
01: 5
Execution
Interval
(seconds)
1: Full Bridge w/mV Excit (P9)
1: 1
Reps
2 25*
5000 mV, 60Hz Rejection, slow Ex Range
3: 22**
50 mV, 60 Hz Rejection, Slow Br Range
4: 1
DIFF
Channel
5: 1
Excite all reps w/Exchan 1
6: 5000
mV
Excitation
7: 1
Loc
[DEPTH_FT]
8:
2.3529 Mult
9: -0.49
Offset
*For 21X use Range Code 5
**For 21X use Range Code 3
; Every 60 minutes output the array ID, time,
; and average water depth in feet
2: If time is (P92)
1: 0000
Minutes
into
a
2: 60
Minute
Interval
3: 10
Set output flag high
3: Real Time (P77)
1: 110
Day, Hour/Minute (midnight = 0000)
4: Average (P71)
1: 1
Reps
2 1
Loc
[DEPTH_FT]