2 edlog – Campbell Scientific CS547A-L Conductivity/Temperature Probe and A547A Interface User Manual

Page 13

Advertising
background image

CS547A Conductivity and Temperature Probe and A547 Interface

'correct errors in the EC measurement due to temperature

Therm107

(TempDeg_C,1,3,Vx2,0,250,1,0)

C25mScm = (Ct * 100)/(((TempDeg_C-25) * TempCoef) + 100)


'end scan loop by calling output table

CallTable

ECSample


NextScan

EndProg

5.2.2 Edlog

*Table 1 Program
01: 5

Execution

Interval

(seconds)


;Make a preliminary measurement of resistance for autoranging.

1: Full Bridge (P6)
1: 1

Reps

2: 15

±2500 mV Fast Range

3: 1

DIFF

Channel

4: 1

Excite all reps w/Exchan 1

5: 2500

mV

Excitation

6: 1

Loc [ Rs ]

7: -.001

Mult

8: 1

Offset


2: BR Transform Rf [X/(1-X)] (P59)
1: 1

Reps

2: 1

Loc [ Rs ]

3: 1

Multiplier

(Rf)

;
;Test the initial measurement to make a more accurate measurement.
;

3: CASE (P93)
1: 1

Case Loc [ Rs ]


4: If Case Location < F (P83)

1: 1.8

F

2: 30

Then

Do


5: AC Half Bridge (P5)

1: 1

Reps

2:

15

±2500 mV Fast Range

3: 2

SE

Channel

4: 1

Excite all reps w/Exchan 1

5: 2500

mV

Excitation

6: 1

Loc [ Rs ]

7: 1.0

Mult

8: 0.0

Offset

7

Advertising