Campbell Scientific CS547A-L Conductivity/Temperature Probe and A547A Interface User Manual
Page 16

CS547A Conductivity and Temperature Probe and A547 Interface
24: IF (X<=>F) (P89)
1: 3
X Loc [ Ct ]
2: 4
<
3: .474
F
4: 30
Then
Do
25: Z=X*F (P37)
1: 3
X Loc [ Ct ]
2: .95031 F
3: 3
Z Loc [ Ct ]
26: Z=X+F (P34)
1: 3
X Loc [ Ct ]
2: -.00378 F
3: 3
Z Loc [ Ct ]
27: Else (P94)
28: Polynomial (P55)
1: 1
Reps
2: 3
X Loc [ Ct ]
3: 3
F(X) Loc [ Ct ]
4: -.02889 C0
5: .98614 C1
6: .02846 C2
7: .000000 C3
8: .000000 C4
9: .000000 C5
29: End (P95)
;This next program set will correct errors in the EC measurement resulting
;from temperature differences.
;
30: Temp (107) (P11)
1: 1
Reps
2: 3
SE
Channel
3: 2
Excite all reps w/E2
4: 4
Loc [ Temp_degC ]
5: 1.0
Mult
6: 0.0
Offset
31: Z=X+F (P34)
1: 4
X Loc [ Temp_degC ]
2: -25
F
3: 6
Z Loc [ A ]
32: Z=X*F (P37)
1: 3
X Loc [ Ct ]
2: 100
F
3: 7
Z Loc [ Ct100 ]
10