Campbell Scientific CS615 Water Content Reflectometer User Manual

Page 14

Advertising
background image

CS615 WATER CONTENT REFLECTOMETER

10

; loop index multiplied by 3 so that 3 readings
measured by pulse
; instruction are advanced 3 locations each pass
through measurement loop

6: Step Loop Index (P90)

1:

3

Step

7: Period Average (SE) (P27)

1:

3

Reps

2:

4

2 V Peak to Peak/200 kHz
Max. Freq.

3:

1

SE Channel

4:

10

No. of Cycles

5:

5

Timeout (units = 0.01 seconds)

6:

1--

Loc [ msec#1 ]

7:

.001

Mult

8:

0.0

Offset

8: End (P95)

; set ports enabling mux and probes low

9: Do (P86)

1:

51

Set Port 1 Low

10: Do (P86)

1:

53

Set Port 3 Low

; apply calibration

11: Polynomial (P55)

1:

48

Reps

2:

1

X Loc [ Period#1 ]

3:

49

F(X) Loc [ WatCont#1 ]

4:

-0.187

C0

5:

0.037

C1

6:

0.335

C2

7:

0.0

C3

8:

0.0

C4

9:

0.0

C5

12: Do (P86)

1:

10

Set Output Flag High

13: Real Time (P77)

1:

10

Hour/Minute

14: Sample (P70)

1:

48

Reps

2:

49

Loc [ WatCont#1 ]

15: End (P95)

End Program