CS615 WATER CONTENT REFLECTOMETER

7

The calibration coefficients in the example
programs are demonstrative only. See Section
9 for information on calibration.

10.1 SAMPLE PROGRAM 1

Simple program using the Period Averaging
Instruction (P27) of a CR10 datalogger to read a
single CS615. The measurement is taken
hourly and the period and water content are
written to output storage.

CR10

CS615

Single-Ended

Channel 1 (SE1)

green

Control Port 5 (C5)

orange

The red lead is connected to 12VDC and the
black and shield are connected to ground.

;{CR10}
;
*Table 1 Program

01:

5.0

Execution Interval (seconds)

1: If time is (P92) ;take reading hourly

1:

0000

Minutes (Seconds --) into a

2:

60

Interval (same units as above)

3:

30

Then Do

2: Do (P86)

1:

45

Set Port 5 High;enable CS615

3: Period Average (SE) (P27)

1:

1

Reps

2:

4

2 V Peak to Peak/200 kHz
Max. Freq.

3:

1

SE Channel

4:

10

No. of Cycles

5:

5

Timeout (units = 0.01 seconds)

6:

1

Loc [ 615period ]

7:

.001

Mult

8:

0.0

Offset

4: Do (P86) ;disable CS615

1:

55

Set Port 5 Low

5: Polynomial (P55) ;convert period to water content

1:

1

Reps

2:

1

X Loc [ 615period ]

3:

2

F(X) Loc [ 615water ]

4:

-0.187

C0

5:

0.037

C1

6:

0.335

C2

7:

0.0

C3

8:

0.0

C4

9:

0.0

C5

6: Do (P86)

1:

10

Set Output Flag High

7: Real Time (P77)

1:

0220

Day,Hour/Minute

8: Sample (P70)

1:

2

Reps

2:

1

Loc [615period]

9: End (P95)

End Program

10.2 SAMPLE PROGRAM 2

Simple program using the Pulse Count Instruction
(P3) of a CR10, CR500 or 21X datalogger to read
a single CS615. The Water Content
Reflectometer is queried when Flag 1 is set high.
The measurement is made and the period and
water content values are written to output storage.

CR10 or 21X

CS615

Pulse Count

Channel 1 (P1)

green

Control Port 5 (C5)

orange

The red lead is connected to 12VDC and the
black and shield are connected to ground.

;{CR10, CR500 or 21X}
;
*Table 1 Program

01:

1.0

Execution Interval (seconds)

1: If Flag/Port (P91) ;set flag1 high to initiate
reading

1:

11

Do if Flag 1 is High

2:

30

Then Do

2: Do (P86) ;enable CS615

1:

45

Set Port 5 High

3: Beginning of Loop (P87) ;delay for complete
count interval

1:

1

Delay

2:

2

Loop Count

4: End (P95)

5: Pulse (P3) ;determine CS615 output frequency

1:

1

Reps

2:

1

Pulse Input Channel

3:

21

Low Level AC, Output Hz

4:

1

Loc [ 615kHz ]

5:

.001

Mult ;convert to kHz

6:

0.0

Offset