Model HFP01SC Self-Calibrating Soil Heat Flux Plate

6.2 Wiring

Connections to Campbell Scientific dataloggers are given in FIGURE 6-1,

TABLE 6-1, and TABLE 6-2. The output of the HFP01SC can be measured

using a single-ended analog measurement (

VoltSE() or Instruction 1),

however, a differential analog measurement (

VoltDiff() or Instruction 2) is

recommended.

The wiring convention is that the white wire is positive with respect to the

green wire, when energy is flowing through the transducer from the side with

the text “this side up” to the other side.

The switched 12 Vdc port can source enough current to calibrate

four HFP01SC plates. If additional HFP01SC plates are needed,

an external relay is required to power the additional plates (see

example 4).

For dataloggers without a SW12V output (CR7X, 21X and CR10), a relay

(A21REL-12) is required for the in-situ calibration (see Example 4).

6.3 Programming

The HFP01SC has a nominal calibration of 15 W m

-2

mV

-1

. Each sensor is

accompanied by a calibration certificate. Each sensor also has a unique

calibration label on it. The label is located on the pigtail end of the sensor

leads.

6.3.1 Example 1. Sample CR3000 Program Using a Differential

Measurement Instruction

TABLE 6-3 provides the wiring for Example 1.

TABLE 6-3. Wiring for Example 1

Description

Color CR3000

Sensor Signal #1

White

9H

Sensor Signal Reference #1

Green

9L

Shield #1

Clear

Sensor Signal #2

White

10H

Sensor Signal Reference #2

Green

10L

Shield #2

Clear

Sensor Signal #3

White

11H

Sensor Signal Reference #3

Green

11L

Shield #3

Clear

Sensor Signal #4

White

12H

Sensor Signal Reference #4

Green

12L

Shield #4

Clear

NOTE