5. normalized response of obs500 to amco clear, Turbidity, High-sediment concentrations – Campbell Scientific OBS500 Smart Turbidity Meter with ClearSensor Technology User Manual
Page 33

OBS500 Smart Turbidity Meter with ClearSensor™ Technology
universal response curve. However, there is residual nonlinearity that is
removed by calibration and by computation of a TU value with a 2nd-order
polynomial. This section explains how to do a turbidity calibration.
0
1000
2000
3000
4000
Turbidity (TU)
0.0
0.2
0.4
0.6
0.8
1.0
N
or
m
al
iz
ed
O
B
S-
3+
R
es
po
ns
e
(O
PV
33
0
VC
SE
L)
FIGURE 7-5. Normalized response of OBS500 to AMCO Clear
®
turbidity. The inset shows the response function of a turbidity
sensor to high-sediment concentrations.
AMCO Clear
®
SDVB turbidity standards are used to calibrate an OBS500
sensor. SDVB standards are made for individual instruments. Standards made
for one model of turbidity meter cannot be used to calibrate a different model.
0
20000
40000
60000
SSC (mg/l)
0
1000
2000
3000
4000
Tu
rb
id
ity
(N
TU
)
A B
C
23