Step 6: Final stage

• Remember to reinstall the plastic blanking screws back into the PWS100

before leaving the site.

• If a CS215-PWS or other piece of equipment was unplugged in order to

perform the tests, then ensure it is reconnected.

• Remember to check that the blanking plug has not been left in the PWS.

• Place all calibration equipment carefully back into its correct box being

careful not to scratch the visibility disk.

Final calibration result table

Result/
LED Status

LED L

LED C

LED R

Description

Test passed

ON

ON

ON

If all tests have passed, all
the LEDs will be illuminated
at the same time

Calibration
required

Flashing Flashing Flashing Calibration

values

are

suspect. Do not use the
calibration values created
from this test

Fault states during tests

If during the calibration cycle, LED C and R start flashing together, then there
has been an error either with the communications to the PWS100 or that the
disk has stalled or is unable to attain the required speed for calibration.

State

LED L

LED C

LED R

Description

Disk

stalled -

Flashing Flashing PWS

calibrator

problem