It is also recommended that the operator stands behind the laser hood while the
PWS is active.

6. Electrical Parameters

6.1 PWC100 Voltages and Currents

Nominal Max

Notes

General details

Standby current

29 mA

30 mA

No tests running

Operating current

125 mA

150 mA

Disk rotating at test
speed

Spin up current

500 mA

700 mA

Disk spinning up to
speed

+ Volts

+12 V

+2 8V

Main supply voltage

6.2 Earth, 0 V and Transient Protection

The user must ensure that the ground lug on the PWS is connected to the earth
lug on the logger or another suitable termination point. The PWC100 is
connected to this earth via the umbilical cable. This connection acts as the
discharge path for electrical transients such as static and surge. Failure to
correctly terminate this connection could lead to damage to the unit or faulty
operation.

The PWC100 is protected against external electrostatic discharges from user
contact but not lightning related surges caused by nearby strikes. Therefore, do
not leave the PWC100 connected to the PWS100 for prolonged periods in field
conditions.