VDIV10.1, VDIV2.1 Voltage Divider 
Terminal Input Modules 
1. Function
Terminal input modules connect directly to the datalogger's input terminals to 
provide completion resistors for resistive bridge measurements, voltage dividers, 
and precision current shunts. Voltage dividers are used to divide a voltage to 
provide a reduced voltage output that is a fraction of the original voltage. The 
VDIV10.1 is a 10:1 voltage divider, the output voltage is one tenth the input 
voltage, and allows a voltage up to ±25 volts to be measured on a ±2500 mV 
range (CR10(X)) and up to ±50 volts to be measured on a ±5000 mV range 
(21X, CR7, CR9000(X), CR23X, CR1000, CR800, CR850, CR3000, CR5000). 
The VDIV2.1 is a 2:1 voltage divider, the output voltage is one half the input 
voltage, and allows a voltage up to ±5 volts to be measured on a ±2500 mV 
range (CR10(X)) and up to ±10 volts to be measured on a ±5000 mV range 
(21X, CR7, CR9000(X), CR23X, CR1000, CR800, CR850, CR3000, CR5000). 
Each voltage divider module may be used to measure one differential voltage 
or two single-ended voltages. 
H
L
AG
H
L
AG
H
L
G
FIGURE 1-1. Terminal Input Module
2. Specifications
2.1 VDIV10.1
10:1 Resistive Divider 
 Resistors 
90 k
Ω/10 kΩ
Ratio Tolerance @ 25
°C
±0.02%
Ratio
Temperature
coefficient
2 ppm/
°C
Power rating
0.25 W
Maximum Input Voltage
50 volts