OPERATING INSTRUCTIONS

The following are basic instructions for operating the PTBT2 ULTRAPEN & App. To

access a full Operations Manual:

1. Inside the App tap the Preferences button on the Feature

Navigation Bar.

2. Select Help from the Settings Menu screen.

3. The browser of your mobile devices will open and link to the Myron L

®

Company

website page where the PTBTx Operations Manual can be accessed.

I. Connecting the ULTRAPEN to the Mobile Device

To pair and connect the PTBT2 to your mobile device:

1. Inside the App tap the Connect button in the Feature Navigation Bar.

2. Push and release the button on the PTBT2 to turn it ON.

3. When the PTBT2’s name appears on the Connect Screen, select that line.

4. A checkmark will appear next to PTBT2’s name.

NOTE: The default name for all PTBT2’s is, “MLC-PTBTx”. To edit the name, Tap the

Preferences button and go to ULTRAPEN Settings section. See the PTBTx

Operations Manual for full instructiions.

To unpair from a PTBT2:

1. Tap the red Unpair button next to the PTBT2’s name.

2. An Unpair Instrument dialogue box will appear.

3. Tap the Disconnect button in the dialogue box.

II. Temperature Unit Selection

The PTBT2 allows you to select either Centigrade (°C) or Farenheit (°F) when displaying

temperature on the iOS screens.

To select the Temperature Units:

1. Inside the App tap the Preferences button on the Feature Navigation Bar.

2. Select Application Preferences from the list of settings topics.

3. Slide the Temperature Units switch left or right to display

temperature values as either °C or °F.

III. pH Mode Selection

The PT2 allows you to select between two pH measurement modes:

Hold (Default): Once stabilized, the reading is captured then displayed.

LIVE: Real-time readings are displayed continuously during measurement until the

ULTRAPEN times out.
To select the pH Mode:

1. Inside the App tap the Preferences button on the Feature

Navigation Bar.

2. Select ULTRAPEN Settings from the list of settings topics

3. Push and release the button on the PTBT2 to turn it ON.

4 When the PTBT2 LED begins to flash, The App will display, “Press and hold the

button on the ULTRAPEN to enter Preferences”

5. Press and hold the PTBT2’s push button.

6. When the LED goes to solid ON the ULTRAPEN Settings screen will appear.

7. Select Pen Mode from the ULTRAPEN Settings screen. The Pen Mode screen will

appear.

8. Tap the Pen Mode field on the screen.

9. Choose the desired mode (Hold or Live) from the list that appears.

10. Tap the Apply button.

IV. pH Measurement

The following table explains the LED Indicator signals and typical durations.

LED Indicator Signal

Meaning

Duration

Rapid Flashing

Dip pen in solution

6 sec

Slow Flashing

Swirl Pen. Measurement in process

In LIVE mode, the readings are

dispayed.

10-30 sec in Hold mode

5 minutes in Live mode

Solid ON Light

Measurement is complete

Final reading is displayed.

≤ 6 sec

To measure solution at the extremes of the specified temperature range,

allow the pen to equilibrate by submerging the cell in the sample solution of the same

temperature for 1 minute prior to taking a measurement.

NOTE: If you cannot dip the pen in the sample solution, pour the sample into a clean

container. If you don’t have a sample container or need to test a vertical stream of

solution, use the scoop.