Myron L PH1 User Manual

Page 9

Advertising
background image

(H ) ions will increase acidity, while an increase in (OH ) ions will increase
alkalinity. The total concentration of ions is fixed as a characteristic of
water, and balance would be 10 mol/liter (H ) and (OH ) ions in a neutral
solution (where pH sensors give 0 voltage).

pH is defined as the negative logarithm of hydrogen ion concentration.
Where (H ) concentration falls below 10 , solutions are less acidic than
neutral, and therefore are alkaline. A concentration of 10 mol/liter of (H )
would have 100 times less (H ) ions than (OH ) ions and be called an
alkaline solution of pH 9.

C. The pH Sensor

The active part of the pH sensor is a thin glass surface which is selectively
receptive to hydrogen ions. Available hydrogen ions in a solution will
accumulate on this surface and a charge will build up across the glass
interface. The voltage can be measured with a very high impedance
voltmeter circuit; the trick is to connect the voltmeter to solution on each
side.

The glass surface encloses a captured solution of potassium chloride
holding an electrode of silver coated with silver chloride. This is as inert a
connection as can be made from metal to an electrolyte. It still can
produce an offset voltage, but using the same materials to connect to the
solution on the other side of the membrane allows the 2 equal offsets to
cancel.

The problem is... the other side of the
membrane is some test solution,
not potassium chloride. The
outside electrode, also called the
Reference Junction, is of the same
construction with a porous plug in
place of a glass barrier to allow the
junction fluid to contact the test
solution without significant
migration of liquids through the
plug material. The figure to the
right shows a typical 2 component
pair. Migration does occur, and this
limits the lifetime of a pH junction from depletion of solution inside the
reference junction or from contamination. The junction is damaged by
drying out because insoluble crystals may form in a layer, obstructing
contact with test solutions. See Cleaning Sensor, pg. 10.

D. The Myron L Integral pH Sensor

The sensor in the pH1 (figure at right)
is a single construction in an easily
replaceable package. The sensor body
holds an oversize solution supply for
long life. The reference junction “wick”
is porous to provide a very stable, low
permeability interface. It is formed in a
ring around the pH sensing electrode.
The construction combines all the best
features of any pH sensor known.

E. Sources of Error

The basics are presented in Cleaning Sensor, pg. 10.

1. Reference Junction

The most common sensor problem will be a clogged junction because a
cell was allowed to dry out. The symptom is a drift in the “zero” setting at 7
pH. This is why the pH1 does not allow more than 1 pH unit of offset
during calibration. At that point the junction is unreliable.

2. Sensitivity Problems

Sensitivity is the receptiveness of the glass surface, which can be
diminished by a film on the surface, or a crack in the glass. These
problems also cause long response time.

3. Temperature Compensation

pH sensor glass changes its sensitivity slightly with temperature, so the
further from pH 7 one is, the more effect will be seen. A pH of 11 at 40 °C
would be off by 0.2 units. The pH1 senses the sensor temperature and
compensates the reading.

14

15

Glass Surface

KCl solution

Junction plug

Electrode wires

Glass
Sleeve

-

+

-

7

+

-

KCl solution

Electrode wire

Glass surface

Junction plug

Electrode wire

H+ ions

-

7

-

9

+

+

+

-

Advertising