Chemglass 30L thru 50L Jacketed User Manual

Page 5

Advertising
background image

www.cglifesciences.com 

 

 

[email protected]

Inlet/Outlet Jacket Connections

The stainless steel adapters are assembled and attached when shipped. Use the following instructions should you need to
remove or adjust these components.

Components Needed for Inlet/Outlet:
2ea CG-1969-A

S.S. Adapters, 1”Beaded Pipe to M16x1.0 Male or M30 x 1.5 Male

2ea CG-1968-66

1”Beaded Pipe Couplings

1. Reactor jacket comes standard with 1”glass beaded pipe on the inlet/outlet.
2. Loosen the nut on the beaded pipe coupling. Wet the PTFE/Viton liner inside

the coupling and attach one side to the 1”beaded pipe on the jacket.

3. Insert the beaded pipe adapter (CG-1969-A) into the other side of the coupling.
4. Tighten the 1”beaded pipe coupling to 35in-lbs using a torque wrench.

MAX JACKET PRESSURE = 12 PSI ON 30L JACKETED REACTOR CG-1968-61

MAX JACKET PRESSURE = 12 PSI ON 50L SHORT JACKETED REACTOR CG-1968-95

MAX JACKET PRESSURE = 12 PSI ON 50L TALL JACKETED REACTOR CG-1968-81

MAX T = 60°C

TEMPERATURE RANGE = -60 to 200°C













The Chemglass High Flow Manifold

System alleviates excessive mechanical

forces on the inlet/outlet and prevents

over pressurizing with a pressure relief

valve. This manifold is available

separately, see CG-1969-M.

Advertising