6 Commissoning, operation

26

WIKA operating instructions digital indicator DI10

6.5.2 Additional setpoints

Menu level Parameter-Ebene




SPCt


Setpoints - Number of additional setpoints, SPCt:

0


In addition to the start and end value, 8 extra setpoints can be
defined to linearize non-linear sensor values. Only the
activated setpoint parameters are displayed.




dIS1


Display values for setpoints dIS1 … dIS8:

8

8

8

8

nOCA

CAL


Under this parameter the value of the setpoints is defined.
With sensor calibration, as with end value/offset, you will be
asked at the end whether a calibration should be made.




InP1


Analogue values for setpoints INP1 … INP8:

0

0

0

0


The setpoints are only displayed with the works calibration (4-
20 mA). Here, the desired analogue values can be freely
selected. The input of constantly rising analogue values must
be carried out by the customer/user.