Gb 3. specifications / 4. design and function, Design and function – WIKA BCS10 User Manual

Page 11

Advertising
background image

11

WIKA operating instructions calibration system, model BCS10

14051046.01 10/2012 GB/D/F/E

GB

3. Specifications / 4. Design and function

Voltage supply

■

Power supply:

3 x 1.5 V AA alkaline batteries

■

Battery life:

approx. 1.500 ... 2.000 h

■

Battery status indication: Icon in display for low battery level

Process connection

■

Test item adapter with quick-acting coupling for G ½

■

Test item adapter with quick-acting coupling for G ¾

■

Test item adapter with quick-acting coupling for M30 x 2

CE conformity

■

EMC directive:

2004/108/EC, EN 61 326-1 emission (group 1, class B)

and interference immunity (appendix A)

Plastic case

■

Dimensions in mm:

395 x 295 x 106

■

Weight:

approx. 4 kg (with contents)

For further specifications see WIKA data sheet SP 60.08 and the order documentation.

4. Design and function

4.1 Short description

The modular calibration system model BCS10 serves for pressure generation and

inspection of mechanical and electronic SF

6

gas density measuring instruments by

means of comparative measurements at 20 °C.

The test pump can generate a pressure of 0 ... 35 bar. Individual measuring points can

be approached with highest accuracy by means of the fine-adjustment valve.

The results of the measurements can be displayed in one of the 21 measuring units or in

customer units using the built-in digital pressure gauge.

4.2 Scope of delivery

■

Precision digital pressure gauge model CPG1000 with attached pneumatic test

pump model CPP30

■

Test item adapter with quick-acting coupling for G ½

■

Test item adapter with quick-acting coupling for G ¾

■

Test item adapter with quick-acting coupling for M30 x 2

■

Adjustable overpressure valve

■

Metal flexible hose for connecting the test item, length 2 m

■

Plastic case, incl. foam insert nad operating instructions

Cross-check scope of delivery with delivery note.

Advertising