SP500 series OEM Programmable Syringe Pump

NE500 series manual

35

Revision 101101

10. Appendix

10.1

Syringe Diameters and Rate Limits

Manufacturer

Syringe

Size

(cc)

Inside

Diameter

(mm)

Maximum

Rate

(ml/hr)

Minimum

Rate

(µl/hr)

B-D

1

4.699

53.07

0.73

3

8.585

177.1

2.434

5

11.99

345.5

4.748

10

14.43

500.4

6.876

20

19.05

872.2

11.99

30

21.59

1120

15.4

60

26.59

1699

23.35

Monoject

1

5.74

79.18

1.088

3

8.941

192.1

2.64

6

12.7

387.6

5.326

12

15.72

593.9

8.161

20

20.12

972.9

13.37

35

23.52

1329

18.27

60

26.64

1705

23.44

Terumo

1

4.7

53.09

0.73

3

8.95

192.5

2.646

5

13

406.1

5.581

10

15.8

600

8.244

20

20.15

975.8

13.41

30

23.1

1282

17.63

60

29.7

2120

29.13

Air-Tite

10

15.9

607.6

8.349

20

20.25

985.5

13.55

30

22.5

1216

16.72

50

29

2021

27.78

Syringe

Size

(uL)

Inside

Diameter

(mm)

Maximum

Rate

(ul/hr)

Minimum

Rate

(µl/hr)

0.5

0.103

25.49

0.001

1

0.146

51.23

0.001

2

0.206

101.9

0.002

Hamilton

Microliter

5

0.326

255.4

0.004

SGE

0.5

0.1

24.03

0.001

1

0.15

54.07

0.001

Advertising