Specifications – A&D Wireless Automatic Blood Pressure Monitor UA851TMW User Manual

Page 20

Advertising
background image

Page E-17 | A&D Medical (Toll-Free): 1-888-726-9966

Specifications

Blood pressure measurements determined by the UA-851THW are

equivalent to those obtained by a trained observer using the cuff/

stethoscope auscultation method within the limits prescribed by the

American National Standards Institute for electronic or automated

sphygmomanometers.

NOTE: Specifications are subject to change without prior notice or any

obligation on the part of the manufacturer.

Model

UA-851THW/UA-851TMW/UA-851THX

Type

Oscillometric

Display

Digital, 20-mm character height

Pressure/pulse displayed simultaneously

Memory

51 readings

Measurement range

Pressure: 20 mmHg to 280 mmHg
Pulse: 40 pulses to 180 pulses

Accuracy

Pressure: ±3 mmHg or ±2%, whichever

is greater
Pulse: ±5%

Pressurization

Automatic, using micropump

Depressurization

Active exhaust valve

Power source

4 type “AA” (1.5 volt) alkaline

batteries (included) or 120 V

AC Adapter (optional)

Battery life

Approximately 4 months with

one daily measurement

Recommended operating

environment

50ºF to 104ºF (10ºC to 40ºC)

Less than 85% relative humidity

Recommended storage environment 14ºF to 140ºF (-10ºC to 60ºC)

Less than 95% relative humidity

Dimensions

Length: 4.7”(120 mm)

Width: 5.9”(150 mm)

Height: 6.1” (156 mm)

Weight

17.6 oz. (500 g) without batteries