Agilent E5250A User’s Guide, Edition 11

10-9

Specifications

Supplemental Data when used with Agilent 4155/4156

Supplemental Data when used with Agilent
4155/4156

The following data shows degradation of Agilent 4155/4156 when used with Low
Leakage I-V Port.

Iout is output current in A, and Vout is output voltage in V.

Supplemental Data when used with Agilent 4155

•

Voltage Range and Offset Accuracy (with MPSMU)

•

Current Range and Offset Accuracy (with MPSMU)

Supplemental Data when used with Agilent 4156 (non-Kelvin connection)

•

Voltage Range and Offset Accuracy (with HRSMU)

•

Current Range and Offset Accuracy (with HRSMU)

Range Setting

Measurement

2 V

960

μV + (0.5 × Iout) V

780

μV + (0.5 × Iout) V

Range Setting

Measurement

1 nA

3 pA + (0.2

× Vout) pA

3 pA + (0.2

× Vout) pA

10 nA

7 pA + (0.2

× Vout) pA

5 pA + (0.2

× Vout) pA

100 nA

50 pA + (0.4

× Vout) pA

30 pA + (0.4

× Vout) pA

Range Setting

Measurement

2 V

500

μV + (0.5 × Iout) V

280

μV + (0.5 × Iout) V

Range Setting

Measurement

10 pA

0.4 pA + (0.2

× Vout) pA

0.1 pA + (0.2

× Vout) pA

100 pA

0.4 pA + (0.2

× Vout) pA

0.1 pA + (0.2

× Vout) pA

1 nA

0.7 pA + (0.2

× Vout) pA

0.5 pA + (0.2

× Vout) pA

10 nA

4 pA + (0.2

× Vout) pA

2 pA + (0.2

× Vout) pA

100 nA

40 pA + (0.3

× Vout) pA

20 pA + (0.3

× Vout) pA