18

Bradyprinter 200M-e and 200M-e 300 User’s Guide

5. Press the (+) key. The front panel display shows “REMOVE RIBBON

CANCEL CONTINUE.”

6. Either remove the ribbon or slide it as far from the printer frame as

possible.

7. Close the printhead, trapping the ribbon in this position.

8. Press the (+) key. The front panel shows “CALIBRATING PLEASE

WAIT.”

9. When this part of the calibration process is completed, the display

reads “RELOAD ALL CONTINUE.”

10. Open the printhead. Pull the backing material until a label is

positioned between the media sensors.

11. Either load the ribbon or return the ribbon to its proper position.

12. Close the printhead. Press the (+) key to perform the next part of the

calibration sequence. “MEDIA AND RIBBON CALIBRATE”
displays. The printer is calibrated when the media stops feeding.

13. Press the SETUP/EXIT key to leave the programming mode. Choose

“permanent” when SAVE CHANGES displays.