Product specifications

Displayed temperature range

93.2 °F -108 °F (34 °C - 42.2 °C)

Operating ambient
temperature range

50 °F -104 °F (10 °C - 40 °C)

Display resolution

0.1 °F or °C

Accuracy for patient temperature range

Maximum Laboratory Error

96.8 °F -102.2 °F (36 °C - 39 °C): ± 0.4 °F

± 0.2 °C

outside this range:

± 0.5 °F

± 0.3 °C

Long term storage ranges
Temperature:

- 4 °F to 122 °F ( - 20 °C to 50 °C)

Humidity:

95 % non-condensing

Battery life:

2 years / 1000 measurements

This infrared thermometer meets requirements established in ASTM
Standard E 1965-98 (for the thermometer system [thermometer with
lens filter]). Full responsibility for the conformance of the product to the
standard is assumed by Braun GmbH, 61476 Kronberg, Germany.

ASTM laboratory accuracy requirements for the thermometer only (not
including a lens filter) in the display range of 98.8 °F to 102.2 °F
(36 °C to 39 °C) for infrared thermometers is ± 0.4 °F (± 0.2 °C), whereas
for mercury-in-glass and electronic thermometers, the requirement per
ASTM Standards E 667-86 and E 1112-86 is ± 0.2 °F ( ± 0.1 °C).

This appliance conforms to the following standards:
DIN EN 60601-1: 3/96 «Medical electrical equipment» –
Part 1: General requirements for safety
DIN EN 12470-5: 2003 «Clinical thermometers» –
Part 5: Performance of infrared ear thermometers
(with maximum device)

Internally Powered Equipment
Continuous Operation
Not Protected against Ingress of Water
U.S. Patent No. 5,088,834 Other Patents Pending

MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For
detailed description of EMC requirements please contact the Braun ThermoScan Customer
Service at 1-800-327-7226.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT.

In the unlikely event you experience any difficulty using the Braun ThermoScan thermometer,
simply call us toll free at 1-800-327-7226.

15

TYPE BF

EQUIPMENT

Attention, consult

ACCOMPANYING

DOCUMENTS

MEDICAL ELECTRICAL EQUIPMENT

CLASSIFIED BY UNDERWRITERS LABORATORIES INC.

®

WITH RESPECT TO ELECTRIC SHOCK, FIRE AND

MECHANICAL HAZARDS ONLY

IN ACCORDANCE WITH UL 2601-1 / CAN CSA C22.2

No.601.1 10HK