Programming, Ethernet connection/communication, Burst trig – BNC 745 User Manual

Page 21

Advertising
background image

Model 745 Digital Delay Generator.

November 2013

8- P

ROGRAMMING

Ethernet connection/communication

Generality

For connection over a LAN, you have to do the following:

• Connect the instrument to the LAN physically,

• In the graphical user interface, specify the LAN address,

• On the control PC, enter the instrument’s IP address,

• After the connection has been established, the following commands can be used to modify

the settings:

o

Set’s the instrument’s IP address with: IP XXX.XXX.XXX.XXX

o

Query’s the instrument’s IP address with: IP? ⇒ :IP XXX.XXX.XXX.XXX

o

Set’s the instrument’s IP mask with: NM XXX.XXX.XXX.XXX

o

Query’s the instrument’s IP mask with: NM? ⇒ :NM XXX.XXX.XXX.XXX

o

Set’s the instrument’s IP gateway: GW XXX.XXX.XXX.XXX

o

Query’s the instrument’s IP gateway with: GW? ⇒ :GW XXX.XXX.XXX.XXX

Command structure

Each command description is composed at least of some of the following items (all commands are

used in a telnet prompt):

• Full command syntax,

• Form Set / Query,

• Brief description,

• Parameters,

• RST value,

• Specified limits.

• Example.

*IDN?

Syntax:

*IDN?

Form:

Query

Description:

Queries instrument identification. Answer gives instrument model,

serial number and firmware version.

Parameter:

-

RST value

-

Example:

BNC,MODEL 745,53901/F,V1.1

Instrument model: MOD745

Serial number: 53901,

Channel number: F => 4 active channels (FF => 8 active channels)

Software version: 1.1

BURST TRIG

Syntax:

BTRIG Source

BTRIG? source

Form:

Set & Query

Description:

Trigger source for Burst mode: internal, external or local

Parameter:

Source: IN1, IN2, EXT, SSL

RST value

-