modulation.

Figure 3-d CW Tab

3.6 Sweeps

Besides the basic CW operation, the instrument supports several sweeping modes. Sweep

parameters can be set in the SWEEP tab (

Figure 3-e

). Select frequency, power or list sweep by

clicking on the FREQ. SWEEP, POWER SWEEP or LIST SWEEP tab respectively.

3.6.1

Frequency Sweep

To start a frequency sweep,

a) enter start and stop frequency in Hz unit

b) enter the number of points (or points per decade)

c) enter the dwell time in milliseconds (min. dwell time is 50 microseconds)

d) enter the off time in microseconds

e) select the sweep mode as linear or log; additionally click “random” if random sequence is

desired

f) select the repetition mode: enter number of repetitions or choose “infinite”

g) select the ALC mode: on hold or off. Generally ALC on is used. If very short dwell times (1-2

ms) are needed, the ALC can be set to “hold”. In this way the ALC loop is disabled during the

sweep. The ALC voltages are calibrated before the sweep to maintain high level accuracy

during the sweep