8 reference in and out, 1 bypass internal 100 mhz reference, Eference – BNC 835 User Manual User Manual

Page 29: N and, Bypass internal 100 mhz reference, Figure 3-o

Advertising
background image


Berkeley Nucleonics Corporation 2955 Kerner Blvd., San Rafael, CA 94901

Phone: 415-453-9955, Fax: 415-453-9956, Email: [email protected], Web: www.berkeleynucleonics.com

Figure 3-n CHIRPS Tab

3.8 Reference In and Out

The internal reference of the signal generator can be locked to an external frequency reference that is

provided to the external reference input on the rear panel. To enable phase lock to the external

reference go to the REFERENCE tab (

Figure 3-o

).

Click Use ext. reference and enter the frequency of the reference that has been externally applied.

The reference frequency must be specified within the frequency range and accurately set to at least 1

ppm. Note that the reference lock will fail if the frequency of the external reference is not accurately

provided.

The lock status is shown as Locked (green) or Not Locked (red).

3.8.1

Bypass internal 100 MHz reference

Some instruments provide the option to bypass (disable) the internal reference and feed an external

100 MHz reference. To enable that operation mode select “External 100 MHz”.

Advertising
This manual is related to the following products: