3 display control, Isplay, Ontrol – BNC 645 User Manual

Page 77

Advertising
background image

77

6. To restore the instrument state to the factory default, press the

softkey under Set to Def. A message will be displayed for you to
confirm the action. Press the Yes softkey to confirm or the Cancel
key to cancel the operation. See Figure 4-19.


Remote interface operation
*SAV {0|1|2|3|4}
*RCL {0|1|2|3|4}
Memory:STATe:NAME {0|1|2|3|4} [,<name>]

5.3 Display Control

Model 645 provides an option to turn off the front display screen. One
may choose to do it for security reason or for speeding up the command
execution from the remote interface. This option is available from the
remote operation only.

Remote interface operation (only)
Use this command to turn off the front panel display:
DISP OFF
Use this command to display a message and turn the display back on:
DISP:TEXT ‘message to display’
Use this command to clear the current message on display.
DISP:TEXT CLEAR


Notes

• The display is automatically enabled when the waveform

generator is powered on, the *RST (reset) command is
issued or it is returned to the local front-panel operation.

• Error messages are always displayed even when the display

is turned off.

• The display setting may be saved with the *SAV command.

When the instrument state is restored using the *RCL
command, the display setting will return to the saved state.


Advertising