BNC 1621M User Manual

Page 26

Advertising
background image

Anisotropy of the device for each energy does not exceed values (in %)
presented in Table 1.1, when the device is rotated in the horizontal plane
(Appendix B, Figure B.1) and values (in %) presented in Table 1.2, when
the device is rotated in the vertical plane (Appendix B, Figure B.2).

25

0.059

0.662

1.25

0

0

0

0

15

±5

±5

±5

30

±10

±10

±10

45

±20

±15

±15

60

±40

±20

±20

-15

±5

±5

±5

-30

±10

±10

±10

-45

±20

±15

±15

-60

±40

±20

±20

Angle of the detection
relative to the direction
of the graduation

Energy of gamma radiation, MeV

Anisotropy

0.059

0.662

1.25

0

0

0

0

15

±5

±5

±5

30

±10

±10

±10

45

±20

±15

±15

60

±40

±20

±20

-15

±5

±5

±5

-30

±10

±10

±10

-45

±20

±15

±15

-60

±40

±20

±20

Angle of the detection
relative to the direction
of the graduation

Energy of gamma radiation MeV

Anisotropy

Table 1.1

Table 1.2

Advertising