SAM 935™ Instruction Manual

7.5

Database Reports

From the U

TIL

menu, it is possible to print a Database Report describing the status of one or all of the

nuclides and their radiation lines. A sample database report is shown here:

This report shows just the Eu152 isotope, but a full database printout would just repeat similar information
for each nuclide in the database. The Enabled column on the left near the nuclide name determines
whether the nuclide as a whole is being processed. The enabled column to the far right shows which radi-
ation lines are enabled for that nuclide. Both must be enabled in order for a line to be processed in the
analysis routines. If the nuclide as a whole is not enabled, individual lines that are enabled will be shown
as (Ena), indicating that they would be enabled if the nuclide was turned on. The efficiency column shows
the efficiencies measured for the standard, or 0.00 if no measurement was ever made. In the latter case,
the efficiency model will be used.

Nuclide Database Printout

Nuclide Halflife Enabled Trig Lvl Energy Intensity Effic SumCorr

Enabled

Eu152 13.542 Y Ena 200U/2k R 41.0 73.80 0.00 4.660 Ena

82.0 0.00 0.00 1.000 Ena

121.8 28.40 0.00 4.600 Ena

162.8 0.00 0.00 1.000 Ena

244.7 7.49 0.00 1.000

344.3 26.60 0.00 1.320 Ena

411.1 2.23 0.00 1.000

444.0 2.78 0.00 1.000

778.9 12.96 0.00 1.210 Ena

867.4 4.15 0.00 1.000

964.1 14.34 0.00 1.000

1101.0 23.68 0.00 1.000 Ena

1408.0 20.85 0.00 1.000 Ena