BNC 1201 - 6 1/2 Digit Digital Multimeter User Manual

Page 122

Advertising
background image

123

accept the trigger. The multimeter takes immediate internal trigger, a

software trigger from a remote interface and external trigger from the

rear panel. Then be sure that the multimeter is ready for a trigger (in the

“wait-for-trigger” state).

The multimeter accepts a trigger only when it is in the “wait-for-trigger”

state. When you have finished configuring the multimeter and have

selected a trigger source, you need to place the multimeter in the

“wait-for-trigger” state so it will take the trigger and make the

measurement. The INITiate, READ? and MEASure? commands all place

the multimeter to the “wait-for-trigger” state.

Triggering Commands

INITiate

READ?

TRIGger:

SOURce {BUS|IMMediate|EXTernal}

SOURce?

TRIGger:

DELay {<seconds>|MINimum|MAXimum}

DELay? [MINimum|MAXimum]

TRIGger:

DELay:AUTO {OFF|ON}

DELay:AUTO?

SAMPle:

COUNt {<value>| MINimum|MAXimum }

COUNt? [MINmum|MAXimum ]

TRIGger:

COUNt {<value>| MINimum|MAXimum|INFinite }

COUNt? [MINmum|MAXimum]