44

Sensor Type

Temperature

Range(°C)

Temperature

Range(°F)

E

-270~1000

-518~1832

J

-210~1200

-140~2192

S

-50~1768

-122~3236

T

-270~400

-518~752

RTD (PT 100)

-200~850

-392~1562

Table 3-1

3.7.1 Thermocouple Measurements

Connect the thermocouple adapter to the banana jacks on the front

panel. The difference between each type is subject to the patch thermal

leads.

※ Note: Only connection via the front panel can be used for temperature

measurements.

How to measure thermocouple

①

Use terminals switch to select terminals on the front panel.

②

To measure thermocouple must be via thermocouple adapter and the

low thermal patch leads as shown in Figure 3-12. (When the

mentioned modules are well connected as the following figure, please

go to the next step.)

③

Configure the thermocouple type and unit using CONFIG + SHIFT +

TEMP, such as K TYPE and ºC. When ready, press ENTER button.

Advertising