Class1 Intelli Tank level display with drip empty User Manual
Page 9

Intelli-Tank Manual (106759)092204
– 9 –
Class 1
The Intelli-Tank display can be calibrated three different ways: 2-point (level
calibration), 5-point and 9-point (volume calibration).
To enter calibration mode use a magnet and activate the magnetic switches in the
order of the appropriate password.
When a magnetic switch is activated the display will indicate which switch was activated
(Left = upper two LEDs, Right = bottom LED) for approximately half a second and then the
display will go blank. There is a limit of 2 seconds between switch activation’s before the
the unit reverts to normal operation.
Entering an invalid password will initiate a “wave off” pattern on the display. (Two center
LEDs, two outer LEDs flashing quickly for 4 cycles.)
Calibrate the unit by entering the desired point calibration –
2 point
RLLR LLRL
5 point
RLLR LRLR
9 point
RLLR RLLR
Calibration mode can be exited at any time by activating the LEFT switch.