IKA ROCKER 3D digital User Manual

Page 15

Advertising
background image

15

Operating modes A and B
Operating mode A

The device does not restart automatically when power is restored after a power outage.

Operating mode B

The device restarts automatically when power is restored after a power outage, using the set values, and displays the (Pr) symbol to
indicate that there has been a power outage within a certain period of time.

Switching between operating modes

To switch between operating modes, switch the device off, then keep the "Start/Stop" button (F) depressed and press the "Power"

button (L).

Adjusting the inclination angle of the shaking table

The inclination angle of the shaking table can be adjusted steplessly from 0 - 15°.

Caution! Do not remove the mat (1) and the cover (2) whilst the device is in operation!

Stop the device by pressing the "Start/Stop" button (F). Do not use your hand to stop the device.

Press the "Power" button (L) to switch the device off.

- To adjust the inclination angle of the shaking table, fi rst remove the mat (1), then the cover (2) that lies beneath it. To do this, press

the recess.

- Slacken the screw (3), using the 4 AF hexagon L-wrench to DIN 911 supplied; prevent the rotary knob from turning.

Caution! Do not fully undo the screw! Slackening the screw half a turn is quite suffi cient to allow adjustment of the inclination angle

of the shaking table.

- Turn the rotary knob to adjust the inclination angle of the shaking table to the desired angle (4).

- Tighten the screw (3) fi nger tight.

- Refi t the cover (2).

3

4

Use of accessories

2

1

Advertising
This manual is related to the following products: