Error message, Accessories – IKA HB 10 User Manual

Page 15

Advertising
background image

15

Error message

Any malfunctions during operation will be identifi ed by an error message on the display (D), see fi g. 5.

Proceed as follows in such cases:

Switch off device using the main switch (B)

Carry out corrective measures

Restart device

Error message Cause

Effect

Correction

E 2

- In remote operation (PC), no communication

Heating off

- Switch off the device

between RV 10 and HB 10

- Clean the IR interface or

- IR connection to RV 10 disconnected

remove objects from around the IR

interface

- Switch on the device

E 3

- Temperature inside the device is too high

Heating off

- Switch off the device

- Let the device cool down

- Switch on the device

E 9

- Fault in storing the safety circuit set

Heating off

- Switch off the device

temperatures

- Let the device cool down

- Memory chip (EPROM) defective

- Switch on the device

E 21

- Safety relay does not open

Heating off

- Switch off the device and switch on again

E 23

- Adjustable safety circuit defective

Heating off

- Switch off the device

- Let the device cool down

- Switch on the device

E 24

- Set safety temperature exceeded

Heating off

- Switch off the device and switch on again

- Let the device cool down

- Switch on the device

- Check the safety temperature

setting

E 25

- The heater control circuit switch (TRIAC) is

Heating off

- Switch off the device and switch on again

faulty. The heater or the supply line is

disconnected.

E 26

- Dry running

Heating off

- Let the device cool down

- Fill up the tempering medium

- Switch off the device and switch on again

E 27

- Error in calibration

Heating off

- Switch off the device and switch on again

E 28

- Breakage of the controller sensor

Heating off

- Switch off the device and switch on again

E 29

- Short-circuit in the safety sensor

Heating off

- Switch off the device and switch on again

E 30

- Short-circuit in the controller sensor

Heating off

- Switch off the device and switch on again

E 31

- Breakage of the safety sensor

Heating off

- Switch off the device and switch on again

E 32

- Temperature deviation is too big

Heating off

- Switch off the device and switch on again

If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- contact the IKA

®

service department,

- send the device for repair, including a short description of the fault.

HB 10.1 Spray guard

Accessories

HB 10.2 Cover (incl. spray guard)

Advertising