Maintenance and cleaning, Technical data – IKA HB 10 User Manual

Page 16

Advertising
background image

16

The device is maintenance-free.

Cleaning

For cleaning disconnect the main
plug.

Only use cleansing agents which have been recommended by
IKA

®

.

To clean the drive use only water with a detergent that contains
tensides, or use isopropylalcohol for stubborn soiling.

Spare parts order

When ordering spare parts, please give:
- Machine type
- Manufacturing number, see type plate
- Item and designation of the spare part,
see www.ika.com, spare parts diagram and spare parts list.

Maintenance and cleaning

Repair

In case of repair the device has to be cleaned and free from any
materials which may constitute a health hazard. For this use the
"certifi cate of compliance" included or download the form from
the IKA

®

website www.ika.com.

If you require servicing, return the appliance in its original
packaging. Storage packaging is not suffi cient. Please also use
suitable transport packaging.

Technical Data

Operating voltage range

Vac

220 - 240 + 10%

100 - 120 + 10%

Rated voltage

Vac

115, 230 - 50/60 Hz

Frequence

Hz

50 / 60

Power input at nominal voltage

W

1350

Fuses

220-240 V T 10 A (5x20)

100-120 V T 16 A (5x20)

Function Heating
Heating output at nominal voltage

W

1350

Temperature range medium

°C

ambient temperature - 180 °C

Temperature indication

LCD-Display

Heating temperature setting

+1

Setting accuracy heating temperature

K

+1

Controller deviation (3l water/ 90 °C)

K

+1

Controller deviation (3l silicone oil/ 50 mPas/

K

+2

150 °C)
Adjustable safety circuit

°C

50-190

Min. volume bath container

l

1

Max. volume bath container

l

3

Materials in contact with medium

stainless steel 1.4404

Safety circuit

°C

adjustable

Dimension (W x D x H)

mm

295 x 265 x190

Weight

kg

2,8

Safety class

DIN 12877, Class 2

Protection acc. to DIN EN 60529

IP 21

Perm. ambient temperature

°C

5 - 40

Perm. humidity

%

80

Protection class

I

Overvoltage category

II

Contamination level

2

Operation at a terrestrial altitude

m

max. 2000 about sea level

Subject to technical changes!