IKA mini G User Manual

Page 19

Advertising
background image

19

Disinfection of surfaces.

- If infectious material escapes into the centrifuge

chamber, disinfect it without delay.

- Use commercially-available disinfectants. Avoid

products that contain aldehydes, phenols, ace-
tone or alcohol (except isopropanol).

- After using a disinfectant, wipe up any residues

using a damp cloth. Surfaces must be dried im-
mediately after being disinfected.

Rotor and accessories

- If the rotor or the accessory parts are contami-

nated with material containing pathogens, ap-
propriate cleaning must be performed.

Autoclaving

The rotor and the accessories may be autoclaved
at 121 °C/ 2 bar.

Centrifuging vessels

- If a centrifuging vessel leaks or breaks, the frag-

ments of the vessel, splinters of glass and the
centrifuged medium that has escaped must be
fully removed.

- If the medium is an infectious material, the cen-

trifuge must be disinfected immediately.

Liquid in the device

Drainage openings are provided in the base of
the centrifuge so that liquids that have entered
inappropriately or condensation can flow out,
preventing damage to the device. If liquid is flow-
ing from the base of the centrifuge disconnect it
immediately from the mains and allow the device
to dry off for a while.

Only clean IKA

®

appliances using these IKA

®

approved cleaning agents. To remove use:
Dyes

Isopropanol (do not

use for the cover)

Building materials

Water containing

detergent,

Isopropanol (do not

use for the cover)

Cosmetics

Water containing

detergent,

Isopropanol (do not

use for the cover)

Food

Water containing

detergent,
Fuels

Water containing

detergent,
Other materials

Please consult IKA

®

Spare parts order

When ordering spare parts, please give:
- Machine type
- Manufacturing number, see type plate
- Item and designation of the spare part,
see www.ika.com, spare parts diagram and

spare parts list

Repair

In case of repair the device has to be cleaned
and free from any materials which may consti-
tute a health hazard. For this use the "certificate
of compliance" included or download the form
from the IKA

®

website www.ika.com.

If you require servicing, return the appliance in
its original packaging. Storage packaging is not
sufficient. Please also use suitable transport pa-
ckaging.

Advertising