• The equipment may heat up by stirring.
• Gradually increase the speed.

Before you fill the reactor vessel, ensure
that the reagents used do not corrode

the FTFE and FKM seal.

• Ensure that the external temperature sensor is inserted in

the media to a depth of at least 20 mm.

• The external temperature sensor must always be inserted

in the media when connected.

• Only use IKA

®

approved accesories!

• Use only original IKA

®

spare parts!

Risk of burns! Exercise caution when
touching the heating block and the glass

vessel. The heating block can reach temperatures in excess
of 230 °C and the glass vessel can reach 120 °C. The cover
of the vessel could reach high temperature when you boil
liquid for long time. Pay attention to the residual heat af-
ter switching off.

• Ensure that the thermostat used for tempering is fully ef-

ficient. A defective thermostat can cause uncontrolledre-
actions.

Only process and heat up any media that
has a flash point higher than the adjust-

ed safe temperature limit that has been set (50...210 ºC).

The safe temperature limit must always be set to at least
25 °C lower than the fire point of the media used.

Use

The IKA

®

LR 1000 Control system is a compact reactor

system. It has been designed to simulate and optimize
chemical reaction processes and can also be used for
mixing, dispersion and homogenization in modeling pro-
cesses.

Operating mode: Tabletop device.

• Range of use (indoor use only)

- Laboratories

- Schools

- Pharmacies

- Universities

Correct use

This equipment is suitable for use in all areas except:
- Residential areas
- Areas that are connected directly to a low-voltage supply

network that also supplies residential areas.

The safety of the user cannot be guaranteed:
- if the instrument is operated with accessories that are not

supplied or recommended by the manufacturer

- if the instrument is operated improperly or contrary to the

manufacturer’s specifications

- if the instrument or the printed circuit board are modified

by third parties.

DANGER

DANGER

WARNING

• Unpacking

- Please unpack the device carefully
- In the case of any damage a detailed report must be

sent immediately (post, rail or forwarder).

• Delivery scope

- LR 1000 control base
- LR 1000 reactor vessel
- LR 2000.61 sensor receptacle
- 2 hose connecting pieces
- Temperature sensor PT 100.5
- USB cable
- Mains cable

Unpacking

- Tools kit

see Fig. 1

- Operating instructions
- Warranty card
- Decontamination clearance certificate.