Accessories

Disperser T 18 digital
Disperser T 25 digital
Disperser adaptor (M36x1.5)
O-ring (Ø 18 x 3.55, FKM)
O-ring (Ø 11.6 x 3, FKM)

Warranty

In accordance with IKA warranty conditions, the warranty

period is 24 months. For claims under the warranty please
contact your local dealer. You may also send the machine
direct to our factory, enclosing the delivery invoice and giv-
ing reasons for the claim. You will be liable for freight costs.

Materials in contact with medium/laboratory reactor

Reactor cover

Stainless steel 1.4571 / RCI 316 L

Reactor cover bottom

Stainless steel 1.4571 / RCI 316 L

Reactor vessel

Borosilicate glass 3.3

Anchor stirrer

Stainless steel 1.4571 / RCI 316 L

Scraper

PTFE / Stainless steel 1.4571 / RCI 316 L

O-ring

FKM

Shaft seal

PTFE

O-ring (Ø 41 x 1.9, FKM)
O-ring (Ø 26 x 2, FKM)
Adaptor for PH meter
Anchor Stirrer with PTFE scraper

The warranty does not cover worn out parts, nor does it
apply to faults resulting from improper use, insufficient care
or maintenance not carried out in accordance with the in-
structions in this operating manual.

Error 17

• SAFE Temp is set at a temperature

lower than the actual heating block
temperature

Heating off • Allow the unit to cool

• Set the SAFE Temp at a higher temperature

Error 26

• Fault in the internal or safety

temperature sensors or measurement
circuit

Heating off • Substitude the internal and safety temperature

sensor.

If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.