Error codes

The fault is shown by an error code on the display (D) as following if the error occurs, e.g. Er 4.

Proceed as follows in such cases:
F

Switch the device switch (B) off.

F

Remove the stirrer tool and remove the instrument from the assembly.

F

Reduce the speed and switch on (instrument switch (A)) the device without the stirrer tool.

Error code

Cause

Effect

Solution

Er 8

Speed sensor fault or overload

motor off

Switch off the instrument

If the actions described fails to resolve the fault or another error code is displayed then take one of the
following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.

Accessories

R 472

Floor stand

R 474

Telescopic stand

R 301

Stirring shaft protection

R 301.1 Support holder

Permitted IKA

®

stirrer tools

Max. Speed rpm

R 1345 Propeller stirrer

≤ 800

R 2302 Propeller stirrer

≤ 600

R 1385 Propeller stirrer

≤ 800

R 1388 Propeller stirrer

≤ 400

R 1302 Dissolver stirrer

≤ 1000

Warranty

In accordance with IKA

®

warranty conditions, the

warranty period is 24 months. For claims under
the warranty please contact your local dealer. You
may also send the machine direct to our factory,
enclosing the delivery invoice and giving reasons
for the claim. You will be liable for freight costs.

Max. Speed rpm

R 1355 Centrifugal stirrer

≤ 800

R 1376 Paddle stirrer

≤ 800

R 2311 Paddle stirrer

≤ 600

R 1333 Anchor stirrer

≤ 800

SI 400 Safety switch

SI 472 Fixing device

SI 474 Fixing device

The warranty does not cover worn out parts, nor
does it apply to faults resulting from improper use,
insufficient care or maintenance not carried out in
accordance with the instructions in this operating
manual.

Advertising