IKA to clean IKA devices. To remove use:
Dyes

isopropyl alcohol

Construction materials

water containing tenside / isopropyl alcohol

Cosmetics

water containing tenside / isopropyl alco-
hol

Foodstuffs

water containing tenside

Fuels

water containing tenside

For materials which are not listed, please request information
from IKA. Wear the proper protective gloves during cleaning of
the devices.
Electrical devices may not be placed in the cleansing agent for
the purpose of cleaning.
Before using another than the recommended method for clea-
ning or decontamination, the user must ascertain with the manu-
facturer that this method does not destroy the instrument.

Thanks to its extensive adjustable range of speeds, its control
and feedback electronics and the exceptional durability of the
device, the IKA Vibrax VXR basic offers a host of options fpr
usage. The available attachments are described below. Please
ask us if require special attachments. All attachment can simply
be placed on the agitation table and fastened in place with the
clamping screws.

VX 1 one-handed attachment
This rubber plate has a soft foam rubber attachment. Several test
tubes can be taken in the hand and pressed onto the foam rub-
ber attachment while the machine is running. It is also possible
to agitate round flasks of up to 100ml. To do this, the round flask
is also pressed onto the moving foam rubber attachment.
Of course it is also possible to place an Erlenmeyer flask or a
Petri dish on the attachment and create a slow vibrating motion
in the medium.

If these conditions are met, the device is ready to operate after
plugging in the mains plug. If these procedures are not followed,
safe operation cannot be guaranteed and/or the equipment may
be damaged.
Please note the ambient conditions indicated in the Technical
details (temperature, and relative humidity).
After the device has been turned on, the control light comes on
three times.

If the motor is locked or if a load is placed on it resulting in a hig-
her temperature than is permitted, the device is turned off auto-
matically by the safety circuit and the control light flashes. To eli-
minate the error, the weight of the material must be reduced.
Allow the device to cool off.
The device must be turned off and backon again.

The IKA Vibrax VXR is maintenace-free. It is subject only to the
natural wear and tear of components and their statistical failure
rate.
When ordering spare parts, please give the manufacturing
number shown on the type plate, the machine type and the
name of the spare part.
If the three silicon hoses that protect the agitation table from
rotating fail, they can easily be replaced using laboratory materi-
al. Use asilicon hose Ø8x12mm and cut it off to alenght of
28,5mm.
Please send in equipment for repair only after it has been clea-
ned and is free from any materials which may constitute a health
hazard. Use only cleansing agents which have been approved by

9

VXR 032001

Motor protection

Accessories

Maintenance and cleaning