Motor protection error codes – IKA MTS 2/4 digital User Manual

Page 11

Advertising
background image

facilitates troubleshooting and helps us draw some initial conclu-
sions.

Error message Error

Cause of error

E 1

The potentiometer for the-Too rapida change in the
speed setting does not

speed target value

specify any target value -Internal device error

E 2

Motor blocked

-Agitation table is inhibi-
ted in its stroke motion by
an external effect

Read fork signal not

-Internal device error

present

E 3

Maximum adjustable

-Resonance behavior

speed of the device

of the test setup

exceeded

E 4

Motor standstill for de-

-To switching on the

vice switch-on routine

device, the motor does

(safety query)

not en rotation motion

not provided

E 5

Safety relay cannot be

-Internal device error

activated

E 7

Problem with power

-The device is being ope-

supply

rated with over-voltage or
under-voltage. (For per-
missible voltage range,
see the Technical Data)

E 8

Triac cannot be activated -Internal device error

The LED display (2) indicates the remaining time in minutes and
counts down backwards until it reaches a value of 0.
When the set time has expired, the drive stops and an audio sig-
nal is heard three times. The LED display (2) shows a value of ”0”
and the dot that identifies Timer mode stops flashing.
You can stop the drive while the timer is running at any time by
pressing the ”Start/Stop” button (4). The LED display (2) will show
a value of ”0” afterward in this case as well, and the dot that iden-
tifies Timer mode will stop flashing.
While Timer mode is in operation, you can increase or decrease
the time setting as required by pressing the ”1” key (6) and/or the
”10” key (5) as described above. After you have changed the time
setting, the time begins to elapse again and the LED display (2)
shows the remaining time in minutes as it did before.

Blocking or overloading the motor beyond the permissible motor
temperature causes the device to switch off automatically due to
the safety circuit. The LED display (2) displays the corresponding
error message. To eliminate the error, please check the operating
conditions. Also allow the device to cool off.
The device must be turned off and back on again.

Error message MTS 2/4:

Please note! If an error occurs, an audio warning signal is gene-
rated and the Timer display switches to E + error number. First try
to turn off the device and then turn it back on to continue operati-
on. If an error is not eliminated even after waiting for a considera-
ble amount of time, please contact our service department. You
should always let us know what error code was displayed. This

­­11

012002

Motor protection

Error codes

Advertising