Product description
Centrifuge 5427 R
English (EN)

14

2.4.6

Rotor F-45-48-5-PCR

2.4.7

Rotor S-24-11-AT

This rotor is only intended for use with 1.5

mL/2.0

mL

tubes.

The following tubes must not be used in this rotor:

• Adapters for 0.2 mL, 0.4 mL, 0.5 mL and 0.6 mL tubes and the corresponding tubes.
• Spin columns

Rotor F-45-48-5-PCR
Fixed-angle rotor for 48 PCR
tubes

Max. g-force:

11

710

×

g

Max. speed:

10

500

rpm

Max. load (tube and
contents):

48

×

0.43

g

Tubes

Tube

Adapter

Adapter base

Max. g-force:

Capacity

Tube diameter

Max. speed

Tubes per adapter/
rotor

Centrifugation radius

conical

11

710

×

g

0.2

mL

Ø

6

mm

10

500

rpm

–/48

9.5

cm

PCR strips

conical

11

710

×

g

0.2

mL

Ø

6

mm

10

500

rpm

–/6

×

8

9.5

cm

Rotor S-24-11-AT

Bucket for tubes:
4

×

1.5

mL/2.0

mL

Max. g-force:

16

049

×

g

Max. speed:

12

700

rpm

Max. load per bucket (tube
and contents):

4

×

3.75

g

Vessels

Vessel

Adapters

Adapter base

Max. g-force:

Capacity

Order no.

(international)

Tube diameter

Max. speed

Vessels per
adapter/rotor

Centrifugation radius

Vessel

16

049

×

g

1.5

mL/2.0

mL

Ø

11

mm

12

700

rpm

-/24

8.9

cm