Product description

Centrifuge 5427 R

English (EN)

9

2.4

Rotors

2.4.1

Rotor FA-45-12-17

Eppendorf centrifuges may only be operated with rotors that are intended for use with the
corresponding centrifuge.

Only use rotors which are marked with the name of the centrifuge (e.g., 5427

R).

Rotor FA-45-12-17
Aerosol-tight fixed-angle rotor
for 12 reaction

tubes

Max. g-force:

20

598

×

g

Max. speed:

14

000

rpm

Max. load (adapter, tube
and contents):

12

×

9.5

g

Vessels

Vessel

Adapters

Adapter base

Max. g-force:

Capacity

Order no.

(international)

Tube diameter

Max. speed

Vessels per
adapter/rotor

Centrifugation radius

Vessel

20

598

×

g

5

mL

Ø

17

mm

14

000

rpm

-/12

9.4

cm

Vessel

5820

768.002

open

17

530

×

g

1.5

mL/2.0

mL

Ø

11

mm

14

000

rpm

1/12

8.0

cm

HPLC vessel

5820

770.007

flat

16

215

×

g

Ø 11

mm

14

000

rpm

1/12

7.4

cm

Cryo tube

5820

769.009

flat

18

188

×

g

1.0

mL

2.0

mL

Ø

13

mm

14

000

rpm

1/12

8.3

cm