25

Operation

Eppendorf BioSpectrometer

®

basic

English (EN)

5.3

Summary of the measuring procedure

5.3.1

Preparing the measurement

1. Switch on the device and, if required, the printer.

The device performs a self test (taking approx. 1 minute) and displays the method selection.

2. Make ready the cuvettes for the measurements (see

Inserting the cuvette on p. 24).

3. Prepare the measuring solutions for measuring the blank values, if required, also the standards and the

samples.

4. Open the cover of the cuvette shaft. The cover can remain open during the measurements.

5.3.2

Measuring procedure

5.3.2.1 Selecting a method

You should not use any measuring solution for standards and samples with a lower
absorbance than 0.02 to 0.03 A (e.g. dsDNA concentration between 1.0 and 1.5 μg/mL). The
detection limit of the device may be significantly lower, nevertheless, the impact of
disturbances from the measuring solutions (particles, bubbles, turbidity) on the reliability of
the result is very high for these low absorbance values.

Use the cursor keys to select the desired
method and call up the method with the

enter

key.

For an overview and a detailed description of the
methods, refer to the next chapter (see

Methods on

p. 31).

Wizard: The wizard at the top of the display will
take you through the method procedure
step-by-step.
Help box: You will receive help texts in the lower
right of the display during each step of the
procedure.
Softkeys: The [< Back] and [Next >] softkeys allow
you to move between method steps in the wizard.