Eppendorf BioSpectrometer basic User Manual
Page 72

Maintenance
Eppendorf BioSpectrometer
®
basic
English (EN)
72
Have the filter recertified by Eppendorf Service after 2 years.
8.3.1.2 Checking wavelength systematic error
To check the wavelength systematic error, proceed as follows: Measure the 3 test filters at the
corresponding wavelength.
4. Results display after measuring a test filter for
testing photometric accuracy.
Measure the other two test filters A2 and A3.
5. Results display after measuring all 3 test filters
for testing the photometric accuracy.
With the and keys, you can view the results
for the different test filters again. Press [Finish] to
complete the test.
6. Compare the average values and CV values to the
supplied table.
If the measured values do not agree with the
permitted range of values, contact Eppendorf
Service.