8 technical data, 1 power supply, 2 ambient conditions – Eppendorf PiezoXpert User Manual
Page 25: 3 weight/dimensions, Eppendorf piezoxpert™ — operating manual
Advertising

Eppendorf PiezoXpert™ — Operating manual
25
8
Technical data
8 Technical data
8.1 Power supply
8.2 Ambient conditions
8.3 Weight/dimensions
Power connection
100 V to 240 V ±10 %, 50 Hz to 60 Hz
Automatically adjusted to the voltage.
Fuses for 100 V to 240 V
T3.15A, 250 V (2 pcs.)
Power consumption
18 W
Protection class
I
Overvoltage category
II
Ambience
Only for use in indoors.
Ambient temperature
15 °C to 35 °C (59 °F to 95 °F)
Relative humidity
30 % to 65 %, dewing not permitted
Atmospheric pressure
80 kPa to 106 kPa
Suitable for use up to a height of 2000 m above sea level
Degree of pollution
2
Weight
2,8 kg (6.17 lb)
Dimensions
Width: 17 cm (6.69 in)
Height: 11.5 cm (4.53 in)
Depth: 23 cm (9.06 in)