Rotors

You can operate the Centrifuge 5424 / 5424 R with the following rotors. Before use of sample
tubes, please note the manufacturer's specifications with regard to centrifugation resistance
(max. rcf).

(1) Maximum load per rotor bore for adapter + tube + content.
(2) Aerosol tightness tested and certified by the Centre for Emergency Preparedness and
Response, Health Protection Agency, Porton Down (UK) (see certificates at the end of this
operating manual).
For the rotors and rotor lids labeled coated, color fluctuations may occur as a result of the
production process. These fluctuations have no effect on service life or resistance to chemicals.

Max. capacity

Max. g-force (rcf) /
speed (rpm)
without adapter
(5424: 120 V,
5424 R)

Max. g-force (rcf) or
speed (rpm)
without adapter
(5424: 230 V/100 V)

Notes

Max. load per rotor bore

(1)

Rotor
FA-45-24-11

24 micro test tubes of 1.5/2.0 mL
each or spin columns.
With adapters:

0.2 mL PCR tubes

0.4 mL/0.5 mL micro test
tubes

0.6 mL Microtainers

21,130 x g /
15,000 rpm

20,238 x g /
14,860 rpm

Aerosol-tight

(2)

rotor

lid (aluminum).

Spin columns
possible, better with
rotor F-45-18-11-Kit.

3.75 g

Rotor
FA-45-24-11-
Special

24 micro test tubes of 1.5/2.0 mL
each or spin columns.
With adapters:

0.2 mL PCR tubes

0.4 mL/0.5 mL micro test
tubes

0.6 mL Microtainers

21,130 x g /
15,000 rpm

20,238 x g /
14,860 rpm

Aerosol-tight

(2)

rotor

lid (aluminum).

PTFE-coated
(particularly resistant
to chemicals),
marked: coated

Spin columns
possible, better with
rotor F-45-18-11-Kit.

3.75 g

Rotor
F-45-18-11-

Kit

18 spin columns or 1.5/2.0 mL
micro test tubes.
With adapters:

0.2 mL PCR tubes

0.4 mL/0.5 mL micro test
tubes

0.6 mL Microtainers

18,111 x g /
15,000 rpm

17,347 x g /
14,860 rpm

Particularly high edge
for all commercially
available spin
columns. See the note
about centrifugation
with open tube lids in
this regard(see Load
the rotor on p. 21)
.

3.75 g

Rotor
F-45-32-5-

PCR

Four PCR strips of 5/8 or 
32 PCR tubes of 0.2 mL each.

18,615 x g /
15,000 rpm

17,829 x g /
14,860 rpm

3.5 g