5 Operation

26

Centrifuge 5424 / 5424 R — Operating manual

5.6.3

Short spin centrifugation

You can carry out a short spin run with the currently set or with the maximum g-force (rcf) or
speed (rpm) of the rotor used. This is set in the device menu (see Device menu on p. 20) before
executing the following steps in the sequence specified:

Performing short spin centrifugation

1. A short run at current g-force (rcf) or speed (rpm) can be set directly using the speed arrow

keys.

2. Only 5424 R: Use the temp arrow keys to adjust the temperature.
3. Start short spin run: Hold down the short key.

In the display flashes while the rotor is running.

The time is counted upwards in seconds.

During short run centrifuging all other keys are blocked.

4. End short spin run: Release the short key.

During the braking process, centrifuging duration flashes in the display.

Only 5424: The centrifuge lid opens automatically. The display shows the symbol .

Only 5424 R: The centrifuge lid remains closed to maintain the sample temperature. You
can open it by pressing the open key.

5. Remove centrifuge contents.

5.6.4

Removing the rotor

1. Turn the rotor nut counterclockwise using the rotor key supplied.
2. Remove rotor by lifting vertically.
3. Only 5424 R: Switch off the centrifuge after use and empty the condensation water tray (pull

out from the left side of the device). Leave centrifuge lid fully opened and protect it against
closing.

Hint!

During the braking process, centrifuging can be restarted up to two more times by pressing the
short key again.

Hint!

Soft ramp is not used for short spin centrifugation.