Technical data, 1 power supply, 2 ambient conditions – Eppendorf 5424 R Centrifuge User Manual

Page 34: Technical data 9.1, Power supply, Ambient conditions, 9 technical data, Oper at ing man u al

Advertising
background image

9 Technical data

Centrifuge 5424 / 5424 R — Operating manual

34

9

Technical data

Operating manual

Centrifuge 5424 / 5424 R

see

on p.

Fig.

Tab.

p.

9.1

Power supply

Centrifuge 5424

Centrifuge 5424 R

9.2

Ambient conditions

Mains power connection

230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz

Current consumption:

1.9 A (230 V)
3.8 A (120 V)
4.0 A (100 V)

Power consumption:

max. 250 W

EMC: Interference emission (radio
interference)

EN 61326 - category B

EMC: Noise immunity

EN 61326

Overvoltage category:

II

Fuses:

3.15 AT (230 V)
6.3 AT (120 V / 100 V)

Mains power connection

230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz

Current consumption:

1.6 A (230 V)
3.2 A (120 V)
3.4 A (100 V)

Power consumption:

max. 350 W

EMC: Interference emission (radio
interference)

EN 61326 - category B

EMC: Noise immunity

EN 61326

Overvoltage category:

II

Fuses:

3.15 AT (230 V)
6.3 AT (120 V / 100 V)

Environment:

For indoor use only.

Ambient temperature:

Centrifuge 5424: 2 to 40°C
Centrifuge 5424 R: 10 to 40°C

Max. relative humidity:

10 to 75 %, non-condensing humidity

Atmospheric pressure:

Use up to an altitude of 2000 m above MSL.

Degree of contamination:

2