7 troubleshooting, 1 general errors, 2 error messages – Eppendorf FemtoJet 4x v.2 User Manual

Page 38: 1 error 01 – 10, Troubleshooting 7.1, General errors, Error messages 7.2.1, Error 01 – 10

Advertising
background image

Troubleshooting
FemtoJet

®

4x

English (EN)

38

7

Troubleshooting

7.1

General errors

7.2

Error messages

7.2.1

Error 01 – 10

Symptom/message

Cause

Remedy

No liquid is injected.

• Capillary is clogged.



Clean the capillary using the

clean

function.



If the error persists, replace
the capillary.

Capillary does not
reach the injection
position.

• The lower safety limit is set

too high.



Set the lower safety limit on
the micromanipulator.

Symptom/message

Cause

Remedy

Error #01

• Software problem



Switch the device off and
back on again.



If the error persists, please
contact the authorized
service.

Error #02

• Compressor control signals

an unexpected error.



Switch the device off and
back on again.



If the error persists, please
contact the authorized
service.

Error #03

• System error



Switch the device off and
back on again.



If the error persists, please
contact the authorized
service.

Error #04

• General electronic error in

the device.



Switch the device off and
back on again.



If the error persists, please
contact the authorized
service.

Error #05

• Faulty hand or foot control.



Switch the device off and
back on again.



If the error persists, please
contact the authorized
service.