5 operation, 5 measuring od 600 – Eppendorf BioPhotometer User Manual
Page 24

66
5 Operation
5.5 Measuring OD 600
Select method
Measuring procedure
Blank measurements remain stored until the date changes
(for further details, see "Measuring nucleic acids").
Measure blank
Measure sample
Measure
next sample
To measure the next sample, press the
key again.
Sample dilution
Sample dilution in the measuring cuvette can be entered
using the
key before the measurement begins and
is then included automatically in the result calculation
(see "Measuring diluted samples").
The OD 600 measurement is a stray-light measurement;
the result is therefore heavily dependent on the geometry
of the light path, which may vary between photometers
from different manufacturers.
5
OD 600
O D 6 0 0
Blank
Sample
or
Blank
O D 6 0 0
B L A N K
0
.
0
0
0
A
Blank
Sample
or
Sample
O D 6 0 0
S A M P L E 0 0 1
0
.
4
3
0
A
Sample
Dilution