6 programming – Eppendorf BioPhotometer User Manual
Page 31

73
6 Programming
The following parameters are offered only when the "Standard" calculation procedure has been programmed:
Parameter
Entries
Explanation
Std. number
Entry of numbers
("1" to "10")
Number of different standards.
Std. measurement
Selection
Selection from "1x", "2x", "3x" repeat measurement of each
standard; a mean value is formed for the further calculation
using the repeat measurements.
Regression
Selection
(Only for standard number of at least 4 (for single
determination of standards: 5))
Selection from the calculation procedure linear and non-linear
regression.
For a number of standards greater than 1 and lower than
4 (or 5 respectively), calculation always takes place via linear
regression (See Section 12, "Calculation").
Std. 1 to Std. 10
Entry of numbers
(five-figure)
Entry of nominal values of standard concentrations;
the number of decimal places of the nominal concentration for
the first standard determines the number of decimal places in
the result.
06_Programm_e.fm Seite 73 Montag, 20. Februar 2006 10:54 Uhr